About the product
The cobas c 303 analytical unit is Roche Diagnostics’ fully automated analyzer for clinical chemistry tests. According to Roche, the instrument is designed for various quantitative and qualitative in vitro tests and contains one photometric measuring unit and one ISE unit. The unit is used as the clinical chemistry and ISE unit in the cobas pure integrated solutions system. For brand information, the Roche Diagnostics page can be consulted.
System features
Photometric measurement is performed with a multi-wavelength spectrophotometer. The ISE unit uses the same sample pipettor. Roche reports that the menu includes more than 110 clinical chemistry applications covering proteins, enzymes, substrates, electrolytes, drugs of abuse, therapeutic drug monitoring and HbA1c tests.
Technical information
- Technology: photometry and ion-selective electrode (ISE), 12 V 50 W lamp, ultrasonic mixer (manufacturer's claim)
- Throughput, photometric only: 450 tests per hour. ISE only: 450 tests per hour (150 samples per hour). Mixed mode: 750 tests per hour (300 photometric and 450 ISE). HbA1c only: 225 tests per hour. Roche states that values may vary with the test mix
- Reagent positions: 42 refrigerated positions, 5 to 15 °C, manual loading, RFID recognition, on-board stability of up to 6 months (varies by test)
- Sample pipetting: 1.0 to 25.0 µL in 0.1 µL steps, 8-second sampling cycle
- Reaction: 128 reaction cuvettes, 75 to 185 µL reaction volume, 37.0 ± 0.1 °C incubation bath, 3 to 10 minute reaction time
- Detection: liquid level, clot and air aspiration detection (available)
- Footprint: 1.2 square meters
Use and compatibility
Roche states that in the cobas pure integrated solutions system the c 303 unit offers 42 refrigerated reagent positions and that c pack green reagents carry up to 6 months of on-board stability within the system. Calibration is performed automatically with cobas AutoCal. The following reagents are listed as compatible with the c 303 on Roche product pages:
- Roche CA2 Gen.2 calcium reagent
- HDLC4 HDL cholesterol reagent
- CK creatine kinase reagent
For system-level information, the cobas pure integrated solutions page can be used.
For product, quotation and technical information requests, the form on the contact page or info@tiliamedical.com can be used.
Frequently asked questions
According to Roche, the unit contains one photometric measuring unit and one ISE unit. The two units use the same sample pipettor.
According to Roche, it runs 450 tests per hour for photometric tests only and 450 tests per hour for ISE tests only. In mixed mode, 750 tests per hour is given. Values may vary with the test mix.
According to Roche, the unit has 42 refrigerated reagent positions. Reagents are loaded manually and recognized by RFID. On-board stability can reach up to 6 months.